Calibration Procedures 2
Keysight DAQ970A/DAQ973A Service Guide 41
To change the security code
To change the security code, you must first unsecure the instrument, and then
enter a new code. Make sure you have read the security code rules described on
page 39 before attempting to change the security code.
Front panel operation:
1 To change the security code, first make sure that the instrument is unsecured.
Press [Utility] > Security.
2 Press New Code, enter the new security code, and press Done.
3 Verify the new code, and press Yes to confirm the change.
4 Press LOCK to secure the instrument with the new code.
Remote interface operation:
CALibration:SECure:CODE <new_code>
Changing the code from the front panel also changes the code as seen from the
remote interface.
All manuals and user guides at all-guides.com
all-guides.com

Table of Contents

Other manuals for Keysight DAQ970A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight DAQ970A and is the answer not in the manual?

Keysight DAQ970A Specifications

General IconGeneral
Scan RateUp to 450 channels/sec
ConnectivityUSB, LAN, GPIB (optional)
Display4.3-inch color LCD
Operating Temperature0 °C to 50 °C
Number of slots3 slots
Measurement TypesDCV, ACV, DCI, ACI, 2-wire resistance, 4-wire resistance, frequency, temperature, capacitance
Voltmeter6.5 digit
Resolution22 bits
MemoryNon-volatile memory