2 Calibration Procedures
46 Keysight DAQ970A/DAQ973A Service Guide
Calibration Actions
Entering calibration values
Use measurement commands to set the instrument's measurement parameters
and functions for the mode being calibrated, then enter calibration values with the
following:
Front panel operation:
[Utility] > Calibrate > Cal Value
After specifying the input value for the calibration step to be performed, press
Perform Cal Step.
Remote interface operation:
Enter cal value: CALibration:VALue <value>
Perform calibration: CALibration[:ALL]?
Storing calibration constants
To store calibration values in non-volatile memory, use the following:
Front panel operation: [Utility] > Calibrate > Save Cal Values
Remote interface operation: CALibration:STORe
All manuals and user guides at all-guides.com
all-guides.com

Table of Contents

Other manuals for Keysight DAQ970A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight DAQ970A and is the answer not in the manual?

Keysight DAQ970A Specifications

General IconGeneral
Scan RateUp to 450 channels/sec
ConnectivityUSB, LAN, GPIB (optional)
Display4.3-inch color LCD
Operating Temperature0 °C to 50 °C
Number of slots3 slots
Measurement TypesDCV, ACV, DCI, ACI, 2-wire resistance, 4-wire resistance, frequency, temperature, capacitance
Voltmeter6.5 digit
Resolution22 bits
MemoryNon-volatile memory