Calibration Procedures 2
Keysight DAQ970A/DAQ973A Service Guide 35
Automating Calibration Procedures
You can automate the complete verification and adjustment procedures outlined
in this chapter if you have access to programmable test equipment. You can
program the instrument configurations specified for each test over the remote
interface. You can then enter readback verification data into a test program and
compare the results to the appropriate test limit values.
You can also adjust the instrument from the remote interface. Remote adjustment
is similar to the local front-panel procedure. You can use a computer to perform
the adjustment by first selecting the required function and range. The calibration
value is sent to the instrument and then the calibration is initiated over the remote
interface. The instrument must be unsecured prior to initiating the calibration
procedure.
For further information on programming the instrument, see the DAQ970A/
DAQ973A Programming Guide.
All manuals and user guides at all-guides.com

Table of Contents

Other manuals for Keysight DAQ970A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight DAQ970A and is the answer not in the manual?

Keysight DAQ970A Specifications

General IconGeneral
Scan RateUp to 450 channels/sec
ConnectivityUSB, LAN, GPIB (optional)
Display4.3-inch color LCD
Operating Temperature0 °C to 50 °C
Number of slots3 slots
Measurement TypesDCV, ACV, DCI, ACI, 2-wire resistance, 4-wire resistance, frequency, temperature, capacitance
Voltmeter6.5 digit
Resolution22 bits
MemoryNon-volatile memory