08 | Keysight | E5071C Network Analyzer - Conguration Guide
1. A sliding load is not supported by the ENA.
2. An offset load is not supported by the ENA.
Measurement Accessories
A complete list of RF and microwave test accessories
is available on our Web site: www.keysight.com/nd/accessories
Accessories are available with the following connector types:
50 Ω Type-N, 3.5 mm, 7 mm, 2.4 mm, 2.92 mm, 1.85 mm, 1.0 mm
and waveguide.
Test port cables and a calibration kit/ECal module should be
added for a complete measurement system.
Calibration kits
Coaxial measurements
Mechanical calibration kits include standards, such as opens,
shorts and loads, which are measured by the network analyzer for
increased measurement accuracy.
Electronic calibration (ECal) kits replace mechanical calibration
standards with one solid-state calibration module that is con-
trolled by the network analyzer via USB. ECal kits provide many
different impedances to the test ports which enables a full two-
port calibration to be performed quickly with a single connection.
This technique reduces operator errors and connector wear and
abrasion.
Choose a calibration kit for each connector type to be used.
Economy, includes:
– open standards (male and female)
– short standards (male and female)
– xed-termination standards (male and female)
Standard, includes the devices in the economy kit and adds:
– sliding load standards1 (male and female) or a series of offset
shorts
Precision, includes the devices in the economy kit and adds:
– 50 Ω airline(s) for TRL calibration
– TRL adapters
Waveguide measurements
For waveguide measurements, Keysight offers mechanical cali-
bration kits that include:
– waveguide-to-coax adapters (X, P, K, R, Q, U, and V bands)
– precision waveguide section
– ush short circuit
– xed terminations
2
– straight section
Cables and adapter sets
Keysight offers the following types of cables:
– single cables: semi-rigid or exible
– cable sets: semi-rigid or exible
There are also adapter sets that protect the test port
and convert the port to the desired connector interface. These
kits contain:
– one male adapter
– one female adapter
To attain the best mechanical rigidity for device connection, use a
single cable and the appropriate special adapter set. To attain the
greatest exibility for device connection, use a cable set.