6 Press the pulse polarity softkey to select positive ( ) or negative ( ) polarity
for the pulse width you want to capture.
The selected pulse polarity is displayed in the upper-right corner of the display.
A positive pulse is higher than the current trigger level or threshold and a
negative pulse is lower than the current trigger level or threshold.
When triggering on a positive pulse, the trigger will occur on the high to low
transition of the pulse if the qualifying condition is true. When triggering on a
negative pulse, the trigger will occur on the low to high transition of the pulse if
the qualifying condition is true.
7 Press the qualifier softkey (< > ><) to select the time qualifier.
The Qualifier softkey can set the oscilloscope to trigger on a pulse width that is:
• Less than a time value (<).
For example, for a positive pulse, if you set t<20 ns:
• Greater than a time value (>).
For example, for a positive pulse, if you set t>20 ns:
• Within a range of time values (><).
For example, for a positive pulse, if you set t>20 ns and t<30 ns:
8 Select the qualifier time set softkey (< or >), then rotate the Entry knob to set
the pulse width qualifier time.
The qualifiers can be set as follows:
• 17 ns to 10 s for > or < qualifier.