Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes
User's Guide
7 Reference Waveforms
To save a waveform to a reference waveform location / 77
To display a reference waveform / 78
To scale and position reference waveforms / 79
To adjust reference waveform skew / 79
To display reference waveform information / 80
To save/recall reference waveform files to/from a USB storage device / 80
Analog channel or math waveforms can be saved to one of two reference
waveform locations in the oscilloscope. Then, a reference waveform can be
displayed and compared against other waveforms. One reference waveform can
be displayed at a time.
You can adjust the vertical scale and offset of reference waveforms. There is also a
skew adjustment for reference waveforms. Reference waveform scale, offset, and
skew information can optionally be included on the oscilloscope display.
Analog channel, math, or reference waveforms can be saved to a reference
waveform file on a USB storage device. You can recall a reference waveform file
from a USB storage device into one of the reference waveform locations.
To save a waveform to a reference waveform location
1 Press the [Analyze] key.
2 Press Features; then, select R1 or R2.
3 Press Features again (or push the Entry knob) to turn on the reference waveform.
4 Press Save/Clear.