24 Keysight M8000 Series of BER Test Solutions Installation Guide
1 System Requirements
 
M8050A Remote Heads 
The M8050A remote heads are connected to M8050A modules, using 
signal and control cables. They allow to generate and probe signals 
directly at the DUT. The M8050A specifications are valid at the end of the 
~150 mm cables of the remote heads. The M8042A pattern generator and 
M8043A error analyzer require a remote head. This means operation 
without their dedicated remote head is not possible.
Remote Heads for M8042A Pattern Generator
The M8058A/M8059A remote head includes an external amplifier that is 
used in combination with the M8042A Pattern Generator. The 64 GBd 
M8058A remote head is used to allow close connection to the device 
under test for symbol rates up to 64 GBd. The 120 GBd M8059A remote 
head is used to allow close connection to the device under test for symbol 
rates up to 120 GBd.
The three cables on the back side of the remote head are used to connect 
with each channel of M8042A Pattern Generator module. 
Figure 12 on 
page -25 shows the M8058A and M8059A remote heads:
Figure 11 M8058A and M8059A Remote Heads for M8042A Pattern Generator
M8058A Pattern Generator  
Remote Head up to 64 GBd
M8059A Pattern Generator  
Remote Head up to 120 GBd