Staircase
Back to ARB Commands
[SOURce:]ARB:CURRent:STAircase:END[:LEVel] <value>, (@<chanlist>)
[SOURce:]ARB:CURRent:STAircase:END[:LEVel]? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:VOLTage:STAircase:END[:LEVel] <value>, (@<chanlist>)
[SOURce:]ARB:VOLTage:STAircase:END[:LEVel]? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:POWer:STAircase:END[:LEVel] <value>, (@<chanlist>)
[SOURce:]ARB:POWer:STAircase:END[:LEVel]? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:RESISTANCE:STAircase:END[:LEVel] <value>, (@<chanlist>)
[SOURce:]ARB:RESISTANCE:STAircase:END[:LEVel]? [MIN|MAX,] (@<chanlist>)
Specifies the ending current, voltage, power, or resistance level after the staircase occurs.
[SOURce:]ARB:SEQuence:STEP:CURRent:STAircase:END[:LEVel] <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:CURRent:STAircase:END[:LEVel]? [MIN|MAX,] <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:STAircase:END[:LEVel] <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:STAircase:END[:LEVel]? [MIN|MAX,] <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:POWer:STAircase:END[:LEVel] <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:POWer:STAircase:END[:LEVel]? [MIN|MAX,] <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:RESISTANCE:STAircase:END[:LEVel] <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:RESISTANCE:STAircase:END[:LEVel]? [MIN|MAX,] <step#>, (@<chanlist>)
Specifies the ending current, voltage, power, or resistance level after the staircase occurs for Arb
sequence steps.
The difference between the start level and the end level is divided equally among the
number of steps.
Parameter Typical Return
Current: 0 - maximum|MIN|MAX, *RST MIN
Voltage: 0 - maximum|MIN|MAX, *RST MIN
Power: 0 - maximum|MIN|MAX, *RST MIN
Resistance: 0 - maximum|MIN|MAX, *RST MIN
Values are range and module-rating dependent.
<end value>
Step 0 - 99, *RST 0 (none)
Programs an end level current: ARB:CURR:STA:END 2,(@1)
Programs an end level voltage: ARB:VOLT:STA:END 20,(@1)
Programs an end level power: ARB:POW:STA:END 20,(@1)
Programs an end level resistance: ARB:RESISTANCE:STA:END 50,(@1)
Programs an end level current for step 0: ARB:SEQ:STEP:CURR:STA:END 2, 0,(@1)
5 SCPI Programming Reference