[SOURce:]ARB:CURRent:STAircase:END:TIMe <value>, (@<chanlist>)
[SOURce:]ARB:CURRent:STAircase:END:TIMe? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:VOLTage:STAircase:END:TIMe<value>, (@<chanlist>)
[SOURce:]ARB:VOLTage:STAircase:END:TIMe? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:POWer:STAircase:END:TIMe<value>, (@<chanlist>)
[SOURce:]ARB:POWer:STAircase:END:TIMe? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:RESISTANCE:STAircase:END:TIMe<value>, (@<chanlist>)
[SOURce:]ARB:RESISTANCE:STAircase:END:TIMe? [MIN|MAX,] (@<chanlist>)
Specifies the time in seconds, after the staircase completes, that the ending current, voltage, power,
or resistance level persists.
[SOURce:]ARB:SEQuence:STEP:CURRent:STAircase:END:TIMe <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:CURRent:STAircase:END:TIMe? [MIN|MAX,] <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:STAircase:END:TIMe <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:STAircase:END:TIMe? [MIN|MAX,] <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:POWer:STAircase:END:TIMe <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:POWer:STAircase:END:TIMe? [MIN|MAX,] <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:RESISTANCE:STAircase:END:TIMe <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:RESISTANCE:STAircase:END:TIMe? [MIN|MAX,] <step#>, (@<chanlist>)
Specifies the time in seconds, after the staircase completes, that the ending current, voltage, power,
or resistance level persists for Arb sequence steps.
Parameter Typical Return
0 262.144 | MIN | MAX, *RST 0 <end time>
Step 0 - 99, *RST 0 (none)
Programs an end time after the current staircase:ARB:CURR:STA:END:TIM 2,(@1)
Programs an end time after the voltage staircase:ARB:VOLT:STA:END:TIM 2,(@1)
Programs an end time after the power staircase:ARB:POW:STA:END:TIM 2,(@1)
Programs an end time after the resistance staircase: ARB:RESISTANCE:STA:END:TIM 2,(@1)
Programs an end time for step 0:ARB:SEQ:STEP:CURR:STA:END:TIM 2, 0,(@1)
[SOURce:]ARB:CURRent:STAircase:NSTeps <value>, (@<chanlist>)
[SOURce:]ARB:CURRent:STAircase:NSTeps? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:VOLTage:STAircase:NSTeps<value>, (@<chanlist>)
[SOURce:]ARB:VOLTage:STAircase:NSTeps? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:POWer:STAircase:NSTeps<value>, (@<chanlist>)
[SOURce:]ARB:POWer:STAircase:NSTeps? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:RESISTANCE:STAircase:NSTeps<value>, (@<chanlist>)
[SOURce:]ARB:RESISTANCE:STAircase:NSTeps? [MIN|MAX,] (@<chanlist>)
Specifies the number of steps in the staircase.
Keysight N6700C Operating and Service Guide 207
5 SCPI Programming Reference

Table of Contents