[SOURce:]VOLTage:RANGe <value>, (@<chanlist>)
[SOURce:]VOLTage:RANGe? [MIN|MAX,] (@<chanlist>)
Sets the output voltage range on models that have multiple ranges. The value that you enter must be
the highest value in volts that you expect to source. The instrument selects the range with the best
resolution for the value entered.
Parameter Typical Return
0 to maximum|MIN|MAX, *RST MAX
The maximum value is 102% of the voltage rating of
the power module.
<voltage range>
Sets the output voltage range to 5 V: VOLT:RANG 5, (@1)
l This command is coupled with [SOURce:]VOLTage. This means that if a range command places an
output on a range with a lower maximum voltage than the present voltage level, an out of range
error occurs. This also occurs if a voltage is programmed that is too large for the present range.
Errors can be avoided by sending both level and range commands in the same SCPI message.
When the range and level information is received as a set, no range/setting conflict occurs.
l If a programmed range change occurs while the output is enabled, the output will be temporarily
disabled while the range switch occurs. The transition from on-to-off and then from off-to-on will
also be delayed by the settings of OUTPut:DELay:FALL and OUTPut:DELay:RISE.
l The query returns the maximum voltage that can be output on the range that is presently set. The
voltage is in the form +n.nnnnnnE+nn for each channel specified. Multiple responses are sep-
arated by commas.
[SOURce:]VOLTage:RESistance[:LEVel][:IMMediate][:AMPLitude] <value>, (@<chanlist>)
[SOURce:]VOLTage:RESistance[:LEVel][:IMMediate][:AMPLitude]?
Sets the output resistance level when in voltage priority mode. Units are in volts.
Parameter Typical Return
0 to 60 V | MIN | MAX, *RST MAX <resistancelevel>
Specifies the voltage priority resistance level: VOLT:RES 30, (@1)
l This command replaces the previous RESistance <value> command and should be used in new
applications. RESistance <value> is still available for backward compatibility.
[SOURce:]VOLTage:RESistance:STATe 0|OFF|1|ON, (@<chanlist>)
[SOURce:]VOLTage:RESistance:STATe? (@<chanlist>)
Enables or disables output resistance in voltage priority mode.
Keysight N6700C Operating and Service Guide 305
5 SCPI Programming Reference