:RELay
:POLarity NORMal|REVerse, (@chanlist) Sets the polarity of the output relays. (Option 760)
:SHORt
[:STATe] 0|OFF|1|ON Simulates a short circuit on the input of the load. (N679xA )
SENSe
:CURRent
:CCOMpensate 0|OFF|1|ON, (@chanlist) Enables/disables the capacitive current compensation. (not on
N678xA SMU, N679xA)
[:DC]
:RANGe
[:UPPer] <value>, (@chanlist) Selects a DC current measurement range.
:AUTO 0|OFF|1|ON, (@chanlist) Enables/disables seamless measurement autoranging. (N678xA SMU)
:ELOG (ELOG commands only on N676xA, N678xA SMU, Option 054)
:CURRent
[:DC]
:RANge
[:UPPer] <value>, (@chanlist) Selects the Elog current measurement range.
:AUTO 0|OFF|1|ON, (@chanlist) Enables/disables seamless measurement autoranging. (N678xA SMU)
:FUNCtion
:CURRent 0|OFF|1|ON, (@chanlist) Enables/disables current data logging.
:MINMax 0|OFF|1|ON, (@chanlist) Enables/disables min/max current data logging.
:VOLTage 0|OFF|1|ON, (@chanlist) Enables/disables voltage data logging.
:MINMax 0|OFF|1|ON, (@chanlist) Enables/disables min/max voltage data logging.
:PERiod <value>, (@chanlist) Sets the integration time of an Elog measurement.
:VOLTage
[:DC]
:RANge
[:UPPer] <value>, (@chanlist) Selects the Elog voltage measurement range.
:AUTO 0|OFF|1|ON, (@chanlist) Enables/disables seamless measurement autoranging. (N678xA SMU)
:FUNCtion <"function">, (@chanlist) Selects the measurement function (for backward compatibility)
:CURRent 0|OFF|1|ON, (@chanlist) Enables/disables current measurements (replaces FUNCtion)
:VOLTage 0|OFF|1|ON, (@chanlist) Enables/disables voltage measurements (replaces FUNCtion)
:INPut MAIN|AUXiliary, (@chanlist) Selects the voltage measurement input. (N6781A, N6785A)
:SWEep (SWEep commands only on N676xA, N678xA SMU, Option 054)
:OFFSet
:POINts <value>, (@chanlist) Defines the offset in a data sweep for triggered measurements.
:POINts <value>, (@chanlist) Defines the number of points in a measurement.
:TINTerval <value>, (@chanlist) Defines the time period between measurement samples.
:RESolution RES20|RES40 Sets the measurement resolution.
:VOLTage
[:DC]
:RANGe
[:UPPer] <value>, (@chanlist) Selects a DC voltage measurement range.
Keysight N6700C Operating and Service Guide 31