Keysight N777xC Series Tunable Laser Family Programming Guide 69
Measurement Operations & Settings 4
378, triggerFreq < min the number of triggers exceeds
the allowed limit
379, Continuous Sweep AND Modulation
= On
379,Continuous Sweep = On
380, LambdaStart < min sweep start is too small
381, LambdaStop > max sweep stop is too large
Example: :sour0:sour0:wav:swe:chec? -> "0,OK"
Affects: All N777xC tunable laser sources
Command: :SOURce0:WAVelength:SWEep:CYCLes
Syntax: :SOURce0:WAVelength:SWEep:CYCLes<wsp><value>|MIN|MAX|DEF|0
Description: Sets the number of cycles.
Cannot be set while a sweep is running.
Parameters: The number of cycles is an integer value.
Also allowed are: MIN: minimum programmable value
MAX: maximum programmable value
DEF: This is not the preset (*RST) default value but is half the sum of, the minimum
programmable value and the maximum programmable value 0: cycles continuously.
Response: None
Example: :sour0:wav:swe:cycl 3
Affects: All N777xC tunable laser sources
Command: :SOURce0:WAVelength:SWEep:CYCLes?
Syntax: :SOURce0:WAVelength:SWEep:CYCLes?[<wsp>MIN|MAX|DEF]
Description: Returns the number of cycles.
Parameters: None
Also allowed are: MIN: minimum programmable value
MAX: maximum programmable value
DEF: This is not the preset (*RST) default value but is half the sum of, the minimum
programmable value and the maximum programmable value 0: cycles continuously.