86 Keysight N777xC Series Tunable Laser Family Programming Guide
4 Measurement Operations & Settings
Triggering - The TRIGger Subsystem
The TRIGger Subsystem allows you to configure how the instrument reacts
to incoming or outgoing triggers.
Command: :TRIGger
Syntax: :TRIGger<wsp>NODEA|1|NODEB|2
Description: Generates a hardware trigger.
Parameters: 1 or NODEA:
2 or NODEB:
Is identical to a trigger at the Input Trigger Connector.
Generates trigger at the Output Trigger Connector.
A hardware trigger cannot be effective in the DISabled triggering mode but can be effective in DEFault, PASSthrough or
LOOPback triggering modes, see :TRIGger:CONFiguration on page 88 for information on triggering modes.
:TRIGger on page 86 describes the :TRIGger command for advanced users.
Response: None
Example: :trig 1
Command: :TRIGger[n]:INPut
Syntax: :TRIGger[n]:INPut<wsp><trigger response>
Description: Sets the incoming trigger response and arms the module.
Parameters: IGNore:
NEXTstep:
SWStart:
Ignore incoming trigger.
Perform next step of a stepped sweep.
Start a sweep cycle.
Response: None
Example: :trig0:inp ign
Affects: All tunable laser modules