EasyManua.ls Logo

Keysight N777 C Series - Trigger[N]:Input; Trigger[N]:Output

Keysight N777 C Series
100 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Keysight N777xC Series Tunable Laser Family Programming Guide 87
Measurement Operations & Settings 4
Command: :TRIGger[n]:INPut?
Syntax: :TRIGger[n]:INPut?
Description: Returns the incoming trigger response.
Parameters: None
Response: IGNore:
NEXTstep:
SWStart:
Ignore incoming trigger.
Perform next step of a stepped sweep.
Start a sweep cycle.
Example: :trig0:inp? -> ign
Affects: All tunable laser modules
Command: :TRIGger[n]:OUTPut
Syntax: :TRIGger[n]:OUTPut
Description: Specifies when an output trigger is generated and arms the module.
Parameters: DISabled:
STFinished:
SWFinished:
SWSTarted:
Never.
When a sweep step finishes.
When sweep cycle finishes.
When a sweep cycle starts.
Response: None
Example: :trig0:outp dis
Affects: All tunable laser modules
Command: :TRIGger[n]:OUTPut?
Syntax: :TRIGger[n]:OUTPut?
Description: Returns the condition that causes an output trigger.
Parameters: none

Table of Contents