Keysight N777xC Series Tunable Laser Family Programming Guide 95
Error Codes 5
-223 “Too much data”
Description:
A function returns more data or the user requests more data than the application is able to handle.
Example:
A tunable laser source produces more data when lambda values of a sweep are stored than the
816x instrument is able to handle. Use the new SENSE:FUNC:RES:BLOCK? command to split the
data aquisition into multiple parts.
-224 “Illegal parameter value”
[Used where exact value, from a list of possibles, was expected.]
-225 "Out of memory"
Description:
The request application or function cannot be executed because the instrument runs out of
memory.
-226 "Lists not same length"
-230 "Data corrupt or stale"
-231 "Data questionable (StatValNYetAcc)"
Description:
The data that is retured is not accurate or reliable. The user should repeat the operation. The
reason for this error is unspecific.
Example:
A powermeter configured a long average time has not completed its current measurement cycle
when the user queries the current power.
-231 "Data questionable (StatRangeTooLow)"
Description:
As -231 (StatValNYetAcc) but for a more specific reason: The powermeter readout data is not
reliable because the currently set (manual) range does not correspond with the input power.
-240 "Hardware error"
-241 "Hardware missing"
-250 "Mass storage error"
-251 "Missing mass storage"
-252 "Missing media"
-253 "Corrupt media"
-254 "Media full"
-255 "Directory full"
-256 "File name not found"