3 VMA Mode
3.9 SEM Measurement
Rel Stop
Sets a relative power level limit at the stop frequency for the selected inner offset,
ranging from –200 to +50 dBc.
The fail condition is set remotely by [:SENSe]:SEMask:OFFSet
[n]:INNer:LIST:TEST for each inner offset channel.
You can turn off (not use) specific inner offset channels remotely with
[:SENSe]:SEMask:OFFSet[n]:INNer:LIST:STATe.
The query returns values currently set to the inner offset stop relative power limits.
When sending the remote command, missing values are not permitted; that is, if you
want to change values 2 and 6, you must send all values up to 6. Subsequent values
remain unchanged. The query for this parameter always returns 12 values.
Remote
Command
[:SENSe]:SEMask:OFFSet[1]|2:INNer:LIST:STOP:RCARrier <rel_ampl>,
[:SENSe]:SEMask:OFFSet[1]|2:INNer:LIST:STOP:RCARrier?
[:SENSe]:SEMask:OFFSet[1]|2:INNer:LIST:STOP:RCARrier:COUPle ON | OFF | 1 | 0,
[:SENSe]:SEMask:OFFSet[1]|2:INNer:LIST:STOP:RCARrier:COUPle?
Example
:SEM:OFFS:INN:LIST:STOP:RCAR -30, -30, -30, -30, -30, -30
:SEM:OFFS:INN:LIST:STOP:RCAR?
:SEM:OFFS:INN:LIST:STOP:RCAR:COUP ON, ON, ON, ON, ON, ON
:SEM:OFFS:INN:LIST:STOP:RCAR:COUP?
Notes Comma-separated list of values
OFFSet 1 is for BTS, 2 for MS. Default is BTS
Couplings Coupled to Rel Start if “Auto” is selected, that is, Start is made the same as Stop
Preset 0dB, 0 dB, 0 dB, 0 dB, 0 dB, 0 dB, 0 dB, 0 dB, 0 dB, 0 dB, 0 dB, 0 dB | 0dB, 0 dB, 0 dB, 0 dB, 0 dB, 0
dB, 0 dB, 0 dB, 0 dB, 0 dB, 0 dB, 0 dB
ON, ON, ON, ON, ON, ON, ON, ON, ON, ON, ON, ON | ON, ON, ON, ON, ON, ON, ON,
ON, ON, ON, ON, ON
State Saved Saved in instrument state
Saved in instrument state
Range Auto|Man
Min/Max -200 dB/50 dB