3 VMA Mode
3.2 Custom OFDM Measurement
3.2.1 Views
The Custom OFDM measurement has four pre-defined views Normal view,
Channel Response view, Demod Error view and Result Summary view.
Some of these are multiple-window Views.
When in a multiple window View, you select a window by touching it. The menu
controls may sometimes change depending on which window is selected.
Whenever the View changes, the default menu is Frequency, unless otherwise
specified in the View description.
View Selection by Enum
Remote
Command
:DISPlay:OFDM:VIEW[:SELect] NORMal | CRESponse | DERRor | NRESults
:DISPlay:OFDM:VIEW[:SELect]?
Example
:DISP:OFDM:VIEW NORM
Sets the normal view
Preset NORM
State Saved Saved in instrument state
Normal
Windows: "IQ Meas Time" on page 415, "Raw Main Time" on page 415, "Spectrum"
on page 415, "Demod Results" on page 418
Multiple windows view consists of the above four windows.
Example
:DISP:OFDM:VIEW NORM
Response
Windows: "CH Freq Response" on page 417, "EQ Impulse Response" on page 417
Dual windows view consists of the above two windows.
Example
:DISP:OFDM:VIEW CRES
Demod Error
Windows: "Error Vector Time" on page 416, "Error Vector Spectrum" on page 416,
"Common Pilot Error" on page 417, "Demod Results" on page 418
Multiple windows view consists of the above four windows.
Example
:DISP:OFDM:VIEW DERR
Vector Modulation Analyzer Mode User's &Programmer's Reference 412