3 VMA Mode
3.4 IQ Waveform Measurement
State Saved Saved in instrument state
Range Marker Noise | Interval Power | Interval Density | Off
Annotation Mkr # <X value> and <Marker value> upper right on graph
Interval Span
Sets the width of the Span for the selected marker.
Remote
Command
:CALCulate:WAVeform:MARKer[1]|2|…|12:FUNCtion:BAND:SPAN <time>
:CALCulate:WAVeform:MARKer[1]|2|…|12:FUNCtion:BAND:SPAN?
Example
:CALC:WAV:MARK:FUNC:BAND:SPAN 20 ms
:CALC:WAV:MARK:FUNC:BAND:SPAN?
Couplings Changing Interval Span necessarily changes "Interval Left" on page 754 and "Interval Right" on page
754
Preset 10% of Meas Time
State Saved Saved in instrument state
Min 0
Max 100 s
Interval Left
Sets the left edge time for the band of the selected marker.
Remote
Command
:CALCulate:WAVeform:MARKer[1]|2|…|12:FUNCtion:BAND:LEFT <time>
:CALCulate:WAVeform:MARKer[1]|2|…|12:FUNCtion:BAND:LEFT?
Example
:CALC:WAV:MARK12:FUNC:BAND:LEFT 1 s
:CALC:WAV:MARK12:FUNC:BAND:LEFT?
Couplings Changing Interval Left necessarily changes "Interval Span" on page 754 and "Interval Right" on page
754
Preset 5% of Meas Time
State Saved Yes
Min 0
Max 100 s
Interval Right
Sets the right edge time for the band of the selected marker.
Remote
:CALCulate:WAVeform:MARKer[1]|2|…|12:FUNCtion:BAND:RIGHt <time>
Vector Modulation Analyzer Mode User's &Programmer's Reference 754