3 VMA Mode
3.2 Custom OFDM Measurement
Remote Command
[:SENSe]:OFDM:AVERage[:STATe] OFF | ON | 0 | 1
[:SENSe]:OFDM:AVERage[:STATe]?
Example
:OFDM:AVER OFF
:OFDM:AVER?
Preset OFF
State Saved Yes
Range Off|On
Average Mode
Toggles the averaging mode between Exp (exponential) and Repeat. This selection
only affects the averaging result after the number of N averages is reached. The
value of N is set using the control “Avg|Hold Number.”
–
EXPonential - Each successive data acquisition after the average count is
reached, is exponentially weighted and then combined with the existing average
–
REPeat - After reaching the average count, averaging is reset and a new average
is started
Remote
Command
[:SENSe]:OFDM:AVERage:TCONtrol EXPonential | REPeat
[:SENSe]:OFDM:AVERage:TCONtrol?
Example
:OFDM:AVER:TCON REP
:OFDM:AVER:TCON?
Notes Selects the type of termination control used for averaging. This determines the averaging action after
the specified number of frames (average count) is reached
Preset EXPonential
State Saved Yes
Range EXPonential|REPeat
Copy CC To
Copies Custom OFDM demodulation parameters from the selected Component
Carrier to another Component Carrier (CC0 – CC7), or to all Component Carriers
(ALL).
Remote
Command
[:SENSe]:OFDM:CCARrier0|…|7:COPY CC0 | CC1 | CC2 | CC3 | CC4 | CC5 | CC6 |
CC7 | ALL
Example
:OFDM:CCAR0:COPY ALL
Couplings Copies the parameters settings of the selected Component Carrier to the target Component Carrier
Vector Modulation Analyzer Mode User's &Programmer's Reference 490