EasyManua.ls Logo

Keysight N9020B - Page 187

Keysight N9020B
2650 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3 VMA Mode
3.1 Digital Demod Measurement
A Average Averaged value in average cycle
P Peak Hold Detected Peak/Maximum value in average cycle
- n/a
The FETCh, MEASure, or READ queries return the following comma-separated scalar
results for Segment SI. For values of SI, see "Index Sub-Fields" on page 184.
For details of available formats, see "Modulation Format" on page 335.
Modulation Format: QPSK
# Name A/P Description Unit, if
any
1 Transmit
Power
A Transmit power
Floating point number
dBm
2 RMS EVM A EVM over the entire measurement area
Floating point number
%
3 Peak EVM P Peak EVM in the measurement area
Floating point number
%
4 Peak EVM
Index
P Location of the symbol that has the largest EVM
5 RMS
Magnitude
Error
A Average magnitude error over the entire measurement area
Floating point number
%
6 Peak
Magnitude
Error
P Peak magnitude error in the measurement area
Floating point number
%
7 Peak
Magnitude
Error Index
P Location of the symbol that has the largest magnitude error
8 RMS Phase
Error
A Average phase error over the entire measurement area
Floating point number
°
9 Peak Phase
Error
P Peak phase error in the measurement area
Floating point number
°
10 Peak Phase
Error Index
P Location of the symbol that has the largest phase error
11 Frequency
Error
A Frequency error in the measured signal
Floating point number
Hz
12 Clock Error A Symbol clock error
Floating point number
Hz
13 I/Q Offset A I and Q error (magnitude squared) offset from the origin
Floating point number
dB
187 Vector Modulation Analyzer Mode User's &Programmer's Reference