7 Save/Recall/Print
7.3 Save
Notes The query returns current instrument state of the active mode in IEEE Block data format. The state is in
a machine-readable format only, as follows:
<syst set preamble><state block data>
Where:
<syst set preamble> is the format:
#NMMM
N = number of digits that comprise MMM
MMM = length in bytes of following data
<state block data> is machine readable state data
Example response: #42016<state data>
The state is recalled by sending the :SYST:SET? response data to the instrument. From example
above::SYST:SET #42016<state data>
2428 Vector Modulation Analyzer Mode User's &Programmer's Reference