3 Spectrum Analyzer Mode
3.7 Burst Power Measurement
Zero span measurement Resolution Bandwidth
Zero span measurement Dwell Time
Harmonics (SA Mode only)
Parameters affected by Auto Couple are:
Resolution Bandwidth
Fundamental Frequency
Dwell Time
Range Table Resolution Bandwidths
Range Table Dwell Times
Meas Preset
Returns parameters for this measurement to their factory default settings.
Remote Command
:CONFigure:TXPower
Example
:CONF:TXP
3.7.8.2 Meas Method
Lets you choose and configure the Measurement Method.
Meas Method
There are two options for this parameter:
Above
Threshold
Level
THReshold
This measurement algorithm is used to capture a time record,
and average only those points in the time record that exceed
the user-specified threshold level. No attempt is made to
position the burst, or to calculate/display burst widths. This
can be used to measure continuous signals, or bursted signals
where the Measured Burst Width algorithm is too restrictive
Measured
Burst Width
BWIDth
This measurement algorithm uses the threshold level to
calculate the burst center, and average those points that lie
within a user-specified burst width that is centered upon the
burst. The burst width parameter is described in more detail
below