3 Spectrum Analyzer Mode
3.7 Burst Power Measurement
3.7.9.2 X Scale
Accesses controls that enable you to set the horizontal scale parameters.
Ref Value
Enables you to set the display X Reference Value.
Remote
Command
:DISPlay:TXPower:WINDow[1]:TRACe:X[:SCALe]:RLEVel <time>
:DISPlay:TXPower:WINDow[1]:TRACe:X[:SCALe]:RLEVel?
Example
:DISP:TXP:WIND:TRAC:X:RLEV 1s
:DISP:TXP:WIND:TRAC:X:RLEV?
Couplings If X "Auto Scaling" on page 1350 is ON, this value is automatically determined by the measurement
result. When you set a value manually, X Auto Scaling automatically changes to OFF
Preset 0.000 s
State Saved Saved in instrument state
Min/Max -10.0 s / 10.0s
Annotation <value> s bottom left of graph
Backwards
Compatibility
SCPI
:DISPlay:TXPower:VIEW[1]:WINDow[1]:TRACe:X[:SCALe]:RLEVel
Scale/Div
Enables you to set the display X scale/division value.
Remote
Command
:DISPlay:TXPower:WINDow[1]:TRACe:X[:SCALe]:PDIVision <time>
:DISPlay:TXPower:WINDow[1]:TRACe:X[:SCALe]:PDIVision?
Example
:DISP:TXP:WIND:TRAC:X:PDIV 1ms
:DISP:TXP:WIND:TRAC:X:PDIV?
Couplings If X "Auto Scaling" on page 1350 is ON, this value is automatically determined by the measurement
result. When you set a value manually, X Auto Scaling automatically changes to OFF
Preset 640.0 us
State Saved Saved in instrument state
Min 1.00 ns
Max 1.0 s
Backwards
Compatibility
SCPI
:DISPlay:TXPower:VIEW[1]:WINDow[1]:TRACe:X[:SCALe]:PDIVision
Spectrum Analyzer Mode User's &Programmer's Reference 1349