3 Spectrum Analyzer Mode
3.9 SEM Measurement
1 is used for both BTS and MS
Preset When the max number of offsets is 6:
Mode Values
SA 1, 1, 1, 1, 1, 1
WCDMA 1, 1, 1, 10, 1, 1|1, 1, 1, 1, 1, 1
When the max number of offsets is 12 in these modes, the preset value of Offset G ~ L is the same as
the Offset F value
Mode Values
MSR 1, 1, 1, 1, 1, 1, 1, 1, 1, 1, 1, 1|1, 1, 1, 1, 1, 1, 1, 1, 1, 1, 1, 1
LTEAFDD, LTEATDD, 5G NR 2, 1, 1, 1, 1, 1, 1, 1, 1, 1, 1, 1|2, 2, 1, 1, 1,1, 1, 1, 1, 1, 1, 1
When the max number of offsets is 14in these modes, the preset value of Offset G ~ N is the same as
the Offset F value
Mode Values
WLAN 1, 1, 1, 1, 1, 1, 1, 1, 1, 1, 1, 1, 1, 1
State Saved Yes
Min/Max 1/1000
Backwards
Compatibility
SCPI
[:SENSe]:SEMask:OFFSet[1]|2:LIST:BWIDth:IMULti
Video BW
Changes the instrument post-detection filter.
When sending the remote command, missing values are not permitted; that is, if you
want to change values 2 and 6, you must send all values up to 6. Subsequent values
will remain unchanged. The query for this parameter returns 14 values for WLAN
mode, 12 values for other modes.
Remote
Command
[:SENSe]:SEMask:OFFSet[1]|2[:OUTer]:LIST:BANDwidth:VIDeo <freq>,
[:SENSe]:SEMask:OFFSet[1]|2[:OUTer]:LIST:BANDwidth:VIDeo?
[:SENSe]:SEMask:OFFSet[1]|2[:OUTer]:LIST:BANDwidth:VIDeo:AUTO OFF | ON | 0 |
1,
[:SENSe]:SEMask:OFFSet[1]|2[:OUTer]:LIST:BANDwidth:VIDeo:AUTO?
Example
:SEM:OFFS2:LIST:BAND:VID 3.00 kHz, 3.00 kHz, 3.00 kHz, 100.0 kHz,100.0 kHz,
100.0 kHz
:SEM:OFFS2:LIST:BAND:VID?
:SEM:OFFS2:LIST:BAND:VID:AUTO ON, ON, ON, ON, ON, ON