3 Spectrum Analyzer Mode
3.12 List Sweep Measurement (Remote Only)
State Saved Saved in instrument state
Range Same as the Sweep Time range available in the Swept SA measurement.
Remote Command
[:SENSe]:LIST:TRIGger:SLOPe:POINts?
Example
LIST:TRIG:SLOP:POIN?
Remote Command
Notes
Only one trigger slope selection is allowed.
3.12.3.15 Sequencing
Defines a sequence for stepping through the list. You must use LIST:SEQ or
LIST:SEQ:AUTO before performing a measurement using the READ command.
When lists are not of the same length (except lengths of 0 and 1) an error “Invalid
List length” would be generated at sequencing time.
Remote
Command
[:SENSe]:LIST:SEQuence <value>[,<value>]
[:SENSe]:LIST:SEQuence?
[:SENSe]:LIST:SEQuence:AUTO ON1
[:SENSe]:LIST:SEQuence:AUTO?
Example
LIST:SEQ 1,2,4,3
Remote
Command Notes
LIST:SEQ: defines a sequence for stepping through the list.
LIST:SEQ:AUTO: when on, the sequence is set to 1 through N, where N is the longest list.
Preset Ascending order: 1 through N
Range Depends on the number of frequency points in your list.
Remote Command
[:SENSe]:LIST:SEQuence:POINts?
Example
LIST:SEQ:POIN?
3.12.3.16 Preselector Center State
This command defines a list of Preselector Center request states. When the state is
set to 1 (true), a preselector center will initiate at the frequency setting before the
measurement. If set to 0 (false), no preselector center will occur. A preselector
center state that is set to true will override a preselector adjust offset frequency.
Remote Command
[:SENSe]:LIST:PCENter <state>[, <state>]
[:SENSe]:LIST:PCENter?
Example
LIST:PCEN 0,1,1,1
Spectrum Analyzer Mode User's &Programmer's Reference 1945