3 Spectrum Analyzer Mode
3.2 Swept SA Measurement
On completion of the Short Calibration, the Open and Short calibration
measurements are averaged (power). The picture with prompt is taken off the screen
and a menu with “Store Cal” and “Exit Without Storing Cal” is displayed. When you
press “Store Cal” the resulting trace is stored to Trace 3. If the “Exit Without Storing
Cal” button is pressed, the Open/Short Cal is cancelled. In either case you return to
the Normalize menu.
546 Spectrum Analyzer Mode User's &Programmer's Reference