3 Spectrum Analyzer Mode
3.3 Channel Power Measurement
If neither of the above is true for a given point, then:
If that point in SecondTrace is equal to maxtracevalue, the resultant point is
mintracevalue.
If that point in SecondTrace is equal to mintracevalue, the resultant point is
maxtracevalue.
How trace math is processed
Whenever a trace math function is turned on, or the parameters and/or operands of
an existing trace math function are changed, the destination trace is cleared. After
the trace is cleared, all x-axis values in the trace, and the domain of the trace, are
set to match the X-Axis settings of the first trace operand. When this is complete, a
new sweep is initiated.
The process of acquiring data, processing it using the math and Average/Hold
functions, and presenting it as trace data, consists of several functional blocks, as
shown below: