1 Getting Started
36 Keysight U3401A User’s and Service Guide
DC Voltage
Five ranges: 500.00 mV, 5.00 V, 50.00 V, 500.00 V, 1000.00 V
–Measurement method: Sigma Delta A-to-D converter
Input impedance: 10 MΩ±2% range (typical)
Input protection: 1000 V on all ranges
1 Press .
2 Connect the red and black test leads to the respective input terminals as
shown in Figure 1-11.
3 Probe the test points and read the display. In autoranging mode, the
multimeter automatically selects the appropriate range and the measurement
is displayed.
Figure 1-11 DCV terminal connection and display