EasyManua.ls Logo

Keysight U3401A - Figure 1-13 ACI RMS or DCI (10 A) Terminal Connection and

Keysight U3401A
99 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1 Getting Started
38 Keysight U3401A User’s and Service Guide
Measuring AC (RMS) or DC Current up to 10 A
Measurement ranges:
5.000 A, 10.000 A for DC or AC RMS continuous
12.000 A DC or AC RMS for maximum 30 seconds
Shunt resistance: 0.01 Ω to 100 Ω for 500 μA to 10 A ranges
Input protection: Internal 25 A, 440 V fuse for 10A terminal,
1 Press or .
2 Press or to select the measurement range.
3 Power off the measured circuit.
4 Connect the red and black test leads to the A input terminal as shown in
Figure 1-13.
5 Probe the test points in series with the circuit.
6 Power on the measured circuit and read the display.
Figure 1-13 ACI RMS or DCI (10A) terminal connection and display
+
ACI or DCI
current source

Table of Contents

Related product manuals