Performance Test 4
Keysight U3401A User’s and Service Guide 73
Recommended Test Equipment
The test equipments recommended for the performance verification procedures
are listed below. If the exact instrument is not available, substitute calibration
standards of equivalent accuracy.
A suggested alternate method would be to use the Keysight 3458A 8½ -digit
digital multimeter to measure less accurate yet stable sources. The output value
measured from the source can be entered into the instrument as the target
calibration value.
Table 4-1 Recommended test equipments
Application Recommended equipment Recommended accuracy requirements
Zero calibration Shorting plug—Dual banana plug with copper wire short
between the two terminals
—
DC voltage Fluke 5520A < 1/5 instrument 1 year spec
DC current Fluke 5520A < 1/5 instrument 1 year spec
Resistance Fluke 5520A < 1/5 instrument 1 year spec
AC voltage Fluke 5520A < 1/5 instrument 1 year spec
AC current Fluke 5520A < 1/5 instrument 1 year spec
Frequency Fluke 5520A < 1/5 instrument 1 year spec
Diode Fluke 5520A < 1/5 instrument 1 year spec