Application Tutorial 3
Keysight U3401A User’s and Service Guide 65
Dual Display Operation Examples
This section describes some practical operations when using dual display feature.
Measure DC Voltage and AC Ripple on a Rectification Circuit
A single measurement for both DC voltage and AC ripple can be displayed
through both displays while testing a rectifier circuit.
1 Connect the red and black test leads to the input terminal and probe the test
points as shown in Figure 3-1.
Figure 3-1 Connection terminal when measuring DC voltage and AC ripple
2 Press to select DC voltage measurement in primary display.
3 Press to select AC voltage measurement for secondary display.
4 Press . Use and to select autoranging or manual range for
secondary display.
NOTE
Press to disable the secondary display.
Press to select the suitable range if DCV+ACV ripple is above the scale
of the current range.
AC/DC voltage
source
+