EasyManua.ls Logo

Keysight U3606B - Page 52

Keysight U3606B
123 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1 Calibration Procedures
52 Keysight U3606B Service Guide
CV noise effect (peak-to-peak or RMS noise)
This test measures the peak-to-peak or RMS output voltage in the frequency
range from 20 Hz to 20 MHz.
1 Turn off the instrument and connect a differential amplifier and a fixed resistor
(30 Ω for S1 range, 2.67 Ω for S2 range, or 0.33 Ω for S2m range) between the
front panel (red) and (black) output terminals (see Figure 1-12, “Test
setup for constant voltage noise effect (peak-to-peak and RMS noise)
verification,” on page 35).
2 Connect a 50 Ω feed-thru termination to the differential amplifier output and
an oscilloscope to the 50 Ω feed-thru termination.
3 Turn on the instrument. Press [Voltage] to select the constant voltage mode.
Ensure that appropriate range is selected.
4 Set the output voltage to the full rated value (30 V for S1 range, 8 V for S2
range, or 1 V for S2m range) and current to full scale. Enable the output.
5 Configure the differential amplifier as follows:
a Set to the AC mode (positive and negative) to remove the DC component.
b Set to the differential mode.
c Set the gain to ×10.
d Set the attenuation to 1.
e Set the low-pass filter to 20 MHz bandwidth limit to filter out input signals
containing higher frequencies.
f Set to the zero precision voltage generator.
g Set the input impedance to 1 ΜΩ.
230 V
ac
207 V
ac
253 V
ac
S1 (30 V) ±0.003 V
S2 (8 V) ±0.003 V
S2m (1000 mV) ±0.0003 V
Table 1-14 Constant voltage source effect (line regulation)
verification test (continued)
Input AC line
Low line
voltage limit
High line
voltage limit
Range (parameter) Error from nominal 1-year

Table of Contents

Related product manuals