1 Calibration Procedures
88 Keysight U3606B Service Guide
7 Press to start the adjustment. The word CALib starts flashing to indicate
that the calibration is in progress.
Successful completion of the adjustment is indicated by a short beep and
the display briefly showing PASS.
An adjustment failure is indicated by a long beep, the display showing FAiL
along with a calibration error number. Correct the problem and repeat this
procedure.
8 Repeat step 3 through step 7 for each gain adjustment point shown in
Table 1-30.
9 Verify the capacitance gain adjustments using the “Optional capacitance gain
verification test” on page 62.
Table 1-30 Capacitance gain adjustment
Adjustment item Input capacitance
OPEn
Input terminals open (remove any test leads or shorting
plugs from the input terminals)
0.400 nF 0.4 nF
1.000 nF 1 nF
10.00 nF 10 nF
100.0 nF 100 nF
1.000 μF1 μF
10.00 μF 10 μF
100.0 μF 100 μF
1000 μF 1000 μF
10000 μF 10000 μF