1.5.3 Defrosting
The unit defrosts automatically every 12 hours.
Automatic defrosting is time and temperature-controlled.
During automatic defrosting, the system ensures that the unit main-
tains the target temperature value.
When the defrosting process is active, the ‘Defrosting’ LED display
lights up on the Display.
In addition to automatic defrosting, it is possible to start the
defrosting process manually.
Press button [Additional defrosting] for four seconds.
ð
The defrosting process begins.
1.5.4 Temperature display
The temperature is shown on the Display of the unit’s display and
control unit
Ä
Chapter 1.2 ‘Display and control elements’
on page 11.
The Display indicates the temperature of the chilled goods.
The temperature display does not indicate the current
air temperature of the interior.
The alarm function and the temperature warning limits are based
on the temperature of the chilled goods as shown on the Display.
1.5.5 Temperature memory
The temperature memory records the maximum and minimum tem-
perature values reached during operation.
The temperature memory begins recording when the target tem-
perature is reached or, at the latest, after two hours of operation.
The temperature memory is reset automatically when the unit is
restarted or reset.
The temperature memory can be retrieved and erased manually
Ä
Chapter 7.3 ‘Retrieving/erasing the temperature memory’
on page 44.