13.2 Technical data
FR MED 95
PRO-ACTIVE
MED 100
PRO-ACTIVE
MED 126
PRO-ACTIVE
MED 200
PRO-ACTIVE
MED 288
PRO-ACTIVE/
ULTIMATE*
Capacity in litres 95 95 120 170 280
Temperature setting
approx. in °C
-5 to -25 +2 to +15 +2 to +15 +2 to +15 +2 to +15
Voltage in V 220–240 220–240 220–240 220–240 220–240
Frequency in Hz 50 50/60 50/60 50/60 50
Refrigerant quantity
in grams
42 43 32 37 30
Power consumption
in watts
123 30 105 30 100
Normal consump-
tion in kWh/24
1.33 0.42 0.58 0.53 0.69
Permissible ambient
temperature in °C
+10 to +32 +10 to +38 +10 to +38 +10 to +38 +10 to +38
Specifies the inte-
rior temperature
increase in minutes
at 25°C ambient
temperature in the
event of a power
failure
106 95 71 145 93
Temperature homo-
geneity at 25°C ref-
erence ambient tem-
perature in K.
1.78 1.33
Temperature con-
stancy at reference
ambient tempera-
ture of 25°C in K.
0.33 0.37
External dimensions
including wall dis-
tance (WxDxH) in
cm
54 x 54 x 82 54 x 54 x 82 54 x 53.5 x 82 54 x 54.5 x
122
67 x 72 x 132
Usable dimensions
(WxDxH) in cm
43 x 36 x 34 40.8 x 32 x 34 40.8 x 32 x 59
(usable depth
below 15 cm
less)
38 x 34 x 72 50.4 x 32 x 92
(usable depth
below 15 cm
less)
External dimensions
with door open 90°
(WxD) in cm
54 x 105.5 54 x 105.5 54 x 105.5 54 x 105.5 67 x 130
Appendix
Technical data
23.05.2023 Refrigerator/freezer MED series 73