Recommended temperature warning limits for
FROSTER
To ensure that the unit functions properly, the tempera-
ture warning limits must be set for FROSTER models
as shown in table
Ä
‘Temperature warning limits’
on page 40.
Tab. 6: Temperature warning limits for MED and FROSTER MED
Model Lower temperature
warning limit
Target temperature Upper temperature
warning limit
MED 100 PRO-ACTIVE
MED 126 PRO-ACTIVE
MED 200 PRO-ACTIVE
MED 288 PRO-ACTIVE
MED 340 PRO-ACTIVE
MED 468 PRO-ACTIVE
MED 520 PRO-ACTIVE
MED 520 CR PRO-ACTIVE
MED 600 PRO-ACTIVE
MED 720 PRO-ACTIVE
MED 288 ULTIMATE
MED 340 ULTIMATE
MED 468 ULTIMATE
MED 520 ULTIMATE
MED 520 CR ULTIMATE
MED 720 ULTIMATE
+2 °C +5 °C +8 °C
FROSTER MED 95 PRO-ACTIVE -35 °C -20 °C -10 °C
6.2.2.2 Displaying and changing the temperature warning limits
1. Press button [Temperature warning max.].
ð
The upper temperature warning limit is shown on the Dis-
play.
Temperature warning limits
Displaying the temperature warning
limits
Upper temperature warning limit
Commissioning
Programming the unit > Temperature warning limits
23.05.2023Refrigerator/freezer MED series40