Diaphragm pump N838_E
Technical data
Translation of Original Operating and Installation Instruction, KNF 347394-347605
09/23
13
Electrical parameters
Parameter Value
Voltage/frequency See type plate
Maximum current draw
Power consumption of the
pump
Protection class of motor:
N838ANE IP 00
N838KNE IP 00
N838.1.2KNE IP 20
Maximum supply voltage fluc-
tuations
± 10%
Tab.8: Electrical parameters
Thermal switch
NOTICE
Pump overload protection
The pumps are protected against over-
load by a thermal switch as standard.
WARNING
Personal injury and damage to the
pump through automatic start
If pump operation is interrupted by the
thermal switch due to overheating, the
pumps start automatically after they
have cooled.
à Ensure that no dangerous situa-
tions can arise as a result.
Weight
Pump type Value
N838ANE / N838KNE 2.6 kg
N838.1.2KNE 4.4 kg
Tab.9: Weight