Diaphragm pump N838_E
Safety
Translation of Original Operating and Installation Instruction, KNF 347394-347605
09/23
9
Handling of com-
bustible media
Note that the pump is not designed to be explosion-proof.
Make certain that the temperature of the medium is always
sufficiently below the ignition temperature of the medium so
as to prevent ignition or explosion. This also applies for ab-
normal operating situations.
Note that the temperature of the medium increases when the
pump compresses the medium.
Therefore, make certain that the temperature of the medium
also remains sufficiently below the ignition temperature of the
medium even when it is compressed to the maximum permis-
sible operating pressure of the pump. The maximum permissi-
ble operating pressure of the pump is stated in Chapter 3
Technical data [}12].
Make certain that the permissible ambient temperature (see 3
Technical data [}12]) is not exceeded.
Where applicable, also take into account external energy
sources (such as radiated heat sources) that could addition-
ally heat the medium.
In case of doubt, contact KNF Customer Service.
2.6 Use
2.6.1 Proper use
The pumps are intended exclusively for transferring gases
and vapors.
The pumps are intended exclusively for operation in indoor ar-
eas and in non-explosive atmospheres.
2.6.2 Foreseeable misuse
The pumps must not be operated in explosive atmospheres.
The pumps are not suitable for transferring the following:
§
Dusts
§
Liquids
§
Aerosols
§
Biological and microbiological substances
EN
-
US
EN