EasyManua.ls Logo

KNF SC 920G - Technical Specifications; Vacuum Pump System Specs

KNF SC 920G
48 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Technical Data Vacuum Pump System SC 920G
8 Translation of original Operating Instructions, English, KNF 121308-121311 07/18
4. Technical Data
4.1. Vacuum pump system
Materials for parts contacting the medium
Pump head
PPS
Diaphragms
PTFE-coated
Valves
FFPM
Sealings
FPM
Hose connectors
PVDF
Tubing
Norprene
®
Pressure sensor
Ceramics
Venting valve
FFPM
Valve block
PP
Gas ballast valve
PVDF
Pneumatic performance
Max. permissible operating
pressure [bar g]
0
Ultimate vacuum [mbar abs.]
≤ 2
≤ 5 with open gas ballast
Delivery rate at atm. pressure
[l/min]*
0-21 (controllable)
Max. permissible pressure at
inert gas connection [bar g]
0.1
Ambient and media temperature
Permissible
ambient temperature
+ 10 °C to + 40 °C
Permissible media temperature
+ 10 °C to + 40 °C
Other parameters
Vacuum connection and gas
outlet
For tube ID 10 mm
Ventilation connection
For tube ID 6 mm
Weight of vacuum pump
system, including hand
terminal [kg]
15.2
Dimensions:
width x height x length [mm]
366 x 423 x 294
Maximum permissible ambient
relative humidity
80 % for temperatures up to
31°C, decreasing linearly to
50 % at 40°C
Maximum altitude of site
[m above sea level]
2000
Tab. 2 (part 1) *Liters in standard state (1,013 mbar at 0 °C)

Related product manuals