EasyManua.ls Logo

Knick Protos 34009(X) - Page 13

Knick Protos 34009(X)
60 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
13
SW 3400-002: Specifiable Buffer Set
Select menu: Parameter setting/System control/Buffer table
Individual buffer set (with 3 buffer solutions) for pH measurement
Buffer table
You can enter an individual buffer set (with 3 buffer solutions). To do so, you
enter the nominal buffer values for the correct temperature (range 0 ... 95 °C /
5°C steps). Then this buffer set is available in addition to the permanently set
standard buffer solutions in the “Calimatic buffer menu (select Table”).
Menu Display
Buffer table: Entering values
To enter a buffer set
Open parameter setting
System control
Select “Buffer table
Select buffer to be entered
Enter nominal buffer value and all
other values for the correct temper-
ature (right/left arrow keys to select
position, up/down arrow keys to
edit number, confirm by pressing
enter.)
The special buffer set is selected as follows:
Parameter setting/Module PH/Cal preset values/Calimatic buffer/Table.
Return
25.6 °C
7.20 pH
Point of measurement
Logbook
Concentration table
Release of options
Calculation Blocks
Buffer table
System control (Administrator)
Return
25.6 °C
7.10 pH
Buffer 2
Buffer 3
Buffer 1
Buffer table (Administrator)
Return
25.6 °C
7.10 pH
Nominal buffer value
pH value at 10 °C
pH value at 15 °C
pH value at 05 °C
pH value at 20 °C
pH value at 00 °C
Buffer 1 (Administrator)
+04.00 pH
+04.00 pH
+04.00 pH
+04.00 pH
+04.00 pH
+04.00 pH

Related product manuals