14
Protos 3400(X)
Menu
Display
ServiceScope
SW 3400-004: ServiceScope (pH)
Select menu: Diagnostics/Module PH/ServiceScope
ServiceScope
(Diagnostics/Module PH 3400-xxx)
Displays the noise levels over the
time. Allows distinction of individual
disturbances, periodic and broadband
disturbances, which is helpful for
troubleshooting. An error message is
generated if the noise level exceeds
the failure limit.
ServiceScope
20.2 °C
Return
Fail
Noise
level
7.00 pH
t
ServiceScope
ServiceScope monitors the pH input signal and checks whether it remains in
the control range. An error message is generated if the input signal exceeds the
failure limit.