Parameter Default Selection / Range
Off, Weak, Medium, Strong, On (000 sec... 999 sec)
Input filter
• Pulse suppression
Weak
A Standard, B Trace sensor (with guard)
*
,
C Trace sensor (without guard), Others
NTC 30 kohms, NTC 22 kohms
With guard, Without guard
On, Off
Entry required for Hamilton OXYGOLD: 1.23
for sensor type A, B, C fix
Off, Failure, Maint. request
Sensor data
• Measure in
• Sensor type
• Temperature probe
**
• Sensor
**
• Reference electrode
**
• Membrane correction
• Polarization voltage
• Sensocheck
Liquids
A Standard
NTC 22 kohms
Without guard
Off
1.00
0500 / 0675 mV
Off
%Air
µg/l, mg/l, ppb, ppm
xxxx h (entry)
Cal preset values
• Cal saturation
• Cal concentration
• Cal timer
%AIR
mg/l
0000 h
Auto, Manual (default value 1013 mbars)
Auto, Manual (default value 1013 mbars)
Pressure correction
• Pressure during meas
• Pressure during cal
Auto
Auto
Salinity, Chlorinity, Conductivity
(00.00 g/kg or 0.000 µS/cm, depending on selection)
Salinity correction
• Input
Salinity
* Sensocheck not possible for trace sensor with guard, therefore disabled
** Can only be set with “Sensor type Others” selected
Membrane correction (with Hamilton OXYGOLD sensor)
The sensor is selected in the Parameter setting / Sensor data menu:
Sensor type: Other
When using the Hamilton OXYGOLD oxygen sensor,
you must specify a membrane correction factor of 1.23.