AuditTrail Log
Point of measurement Tank UF0815
FRONT 3400-011 0006123
BASE 3400-021 0006458
PHU 3400-110 0007221
OXY 3400-062 0006045
6653 07/06/09 10:43:02 Login: Admin (ADMIN) OK
6654 07/06/09 10:43:02 Login: > Parameter setting (Administrator) OK
6655 07/06/09 10:43:02 Login: > [1] Module PHU 3400-110 (Admin.) OK
6656 07/06/09 10:43:02 Login: > Parameter setting (Administrator) OK
6657 07/06/09 10:43:02 Login: < Menu selection OK
6658 07/06/09 10:43:02 Login: Admin (ADMIN) OK
6659 07/06/09 10:43:02 Login: > Calibration OK
6660 07/06/09 10:43:02 Login: > [1] Module PHU 3400-110 OK
6661 07/06/09 10:43:02 Login: > [1] Calimatic OK
6662 07/06/09 10:43:02 Login: Warn Identical Buf OK
6663 07/06/09 10:43:02 Login: > [1] Calimatic OK
6664 07/06/09 10:43:02 Login: [1] Cal record OK
6665 Last calibration: 05/31/09 10:04 OK
6666 User: ADMIN OK
6667 Sensor type: El 204 OK
6668 Serial number: 0815 OK
6669 Cal mode: Calimatic OK
6670 Zero point 7.02 pH OK
6671 Slope: 58.0 mV/pH OK
6672 Impedance glass (25°C): 825.9 Mohms OK
6673 Impedance ref (25°C): 119.4 kohms OK
6674 0 1st nominal buffer value: 7.00 pH OK
6675 Setpoint: 6.97 pH OK
6676 Actual value: 7.10 pH OK
6677 Delta value: 0.13 pH OK
6678 0 Electrode potential: 1 mV OK
6679 0 Cal temperature: 25.0 °C OK
6680 0 Response time: 19 s OK
6681 07/06/09 10:43:02 Login: > [1] Module PHU 3400-110 OK
6682 07/06/09 10:43:02 Login: > Calibration OK
Typical AuditTrail log
(gray: logged measurement deviations)