Krautkramer DynaPOCKET Issue 03, 04/2005 3-7
OperationHardness measurement
Closing a measurement set
If you use measurement sets, you can close a mea-
surement set at any time and start a new set. The pre-
vious measured values stored for the calculation and
their displayed arithmetic mean are deleted.
Press the ENTER/SET key. The actual average
value flashes.
Press the ENTER/SET key a second time to close
the current measurement set and to delete the
stored values. The display indicates 0 as a current
reading.
Deleting measured values
If you are using measurement sets, you can delete the
most recent reading at any time to remove an erraneous
measurement from the statistical average.
Press the MODE/CLEAR key for about 2 seconds.
The last value measured flashes.
Press the MODE/CLEAR key a second time to
delete the value. The measured value is deleted and
the average is automatically recalculated.
You can repeat the procedure to sequence through
the readings deleting each until all measured values
are deleted.
H Note:
As long as the display is flashing, you can abort the
deletion.
Briefly press the ENTER/SET key to cancel the
process of deleting.