Krautkramer DynaPOCKET Issue 03, 04/2005 5-3
Error codes Function check and troubleshooting
Error codes
Code Cause Action
E 0.0 Internal error in the EPROM/RAM Remove the batteries and re-insert
them after approx. 1 minute.
Switch the instrument on; if the error
occurs again, contact your nearest
GE Inspection Technologies Service.
E 0.1 Battery voltage is too low Exchange the batteries.
E 0.2 Error with self-test Switch the instrument on again; if the
error occurs again, contact your nearest
GE Inspection Technologies Service.
E 1.0 Evaluation error, impossible to carry out Carry out the measurement once again.
a correct evaluation of the measurement. If necessary, clean the impact device.
E 1.1 Evaluation of the measurement is not yet Please carry out the measurement again.
complete.