9.2.2 Electromagnetic compatibility, immunity
Guidelines and manufacturer's declaration -
Electromagnetic compatibility, immunity to interference
The Kröber O2 is intended for operation in an environment as specified below. The operator of the
Kröber O2 must ensure that it is operated in such an environment.
Electromagnetic
environment - Guidelines
Discharge of static
electricity according
to IEC
61000-4-2
±6 kV Contact
discharge
±8 kV air discharge
±6 kV Contact
discharge
±8 kV air discharge
Floors should be made of
wood or concrete, or have
ceramic tiles. If the floor is
covered with synthetic -
material, the relative
humidity must be at least
30%.
fast transient
electrical
disturbances/bursts
according to IEC
61000-4-4
±2 kV for power lines
±1 kV for input/output
lines
±2 kV for power lines
±1 kV for input/output
lines
The quality of the supply
voltage should be that of a
typical business or hospital
environment.
Surges according
to
IEC 61000-4-5
The quality of the supply
voltage should be that of a
typical business or hospital
environment.
Voltage dips, short-
term interruptions -
and fluctuations in
the supply voltage
IEC 61000-4-11
< 5 % UT
(>95 % dip of UT) for
½ period
The quality of the supply
voltage should be that of a
typical business or hospital
environment.
NOTE:
UT
is the AC line voltage -
before applying the test levels.
40 %
UT
(60 % dip of
UT
) for 5
periods
Power failure alarm
Device restart
70 %
UT
(30 % dip of
UT
) for 25
periods
<5 % UT
(95 % dip of
UT
) for 5 s
Power failure alarm
Device restart
Magnetic field at a
supply frequency -
(50 Hz) according
to
IEC 61000-4-8
Magnetic fields at the
mains frequency should -
correspond to typical
values found in business
and hospital environments.
Portable and mobile radios
should not be used at a -
distance from the Kröber
O2, including lines, less -
than the recommended
separation distance -