Abnormal conditions and their correction
64
7. Abnormal conditions and their correction
7.1 Alarm conditions
Alarm category
/LC display
Description
Temperature
alarm
The operating temperature in the concentrator interior has risen
above 60°C.
Countermeasures:
The device must be switched off immediately.
Check that air can enter the unit unhindered. Furthermore, it
must be ensured that there is sufficient distance to objects (wall,
cabinet etc.) and to heaters etc.
If the coarse dust filter is clogged: replace it.
NOTE!
To protect the patient, oxygen delivery is stopped
immediately. However, the compressor continues to run.
In case of defective fan: replacement
In case of defective compressor: replacement or
maintenance