Abnormal conditions and their correction
65
Alarm category
/LC display
Description
Power failure
alarm
The power supply to the device is interrupted. This leads to an
immediate functional failure of the Kröber O2!
Countermeasures
The following should be checked:
Is the power cord securely plugged into the outlet?
Has a fuse blown? Check fuse and replace if necessary.
Mains switch defective? - Exchange control board
Error in the wiring? - Check wiring with device open
Note
If a functional test of the power failure alarm is to be performed, this
can be done as follows:
-Unplug the power cord from the wall outlet.
-Switch on the device.
The alarm works when activated now during the power-on self-test.
NOTE!
If the alarm is triggered again after switching on, the service
department must be notified.
NOTE!
The power failure alarm is powered by an internal,
maintenance-free capacitor.
Oxygen
deficiency alarm
The Kröber O2 oxygen concentrator is equipped with an innovative
multifunction sensor that monitors the oxygen concentration of the
delivered oxygen.
If this is less than 60%, then an oxygen deficiency alarm is
triggered.
Countermeasures
check humidifier and hoses (internal and external) for
leaks.
defect of the valves: replacement of the functional unit