Kröber O2
Abnormal conditions and their correction
66
Alarm category
/LC display
Description
Oxygen status
alarm
The Kröber O2 oxygen concentrator is equipped with an innovative
multifunction sensor that monitors the oxygen concentration of the
delivered oxygen.
If this is less than 82%, then an oxygen status alarm is triggered.
Countermeasures
Check humidifier and hoses (internal and external) for leaks.
Defect of the valves: replacement of the functional unit
Sensor alarm
There is a malfunction of the multifunction sensor, oxygen quantity
and oxygen concentration cannot be determined with sufficient
accuracy.
Countermeasures
Internal tubing check for kinks, etc.
Check if the accessories are connected correctly.
System alarm
There is a microprocessor error.
Countermeasures
Replacement of the control board.