EasyManua.ls Logo

KROHNE OPTIMASS 8300C - OPTIMASS;OPTIGAS 4000, 4010 C Protection and Marking

KROHNE OPTIMASS 8300C
40 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
11
Power Supply Circuit Data Circuit
Ui = 16.5V Ui = 11.8V
Ii = 340mA Ii = 40mA
Pi = 1.3W Pi = 120mW
Ci = 35nF Ci = 35nF
Li = 10uH Li = 10uH
1.7 OPTIMASS 4000 / 4010C / OPTIGAS 4000 / 4010C
The OPTIMASS 4000 / 4010C OPTIGAS 4000 / 4010C mass flow sensors / mass flow meters are
designed with intrinsically safe protection type. The marking for the OPTIMASS 4000 / 4010C /
OPTIGAS 4000 / 4010C is as follows:
II 1/2 G Ex ib IIC T4...T1 Ga/Gb
II 2 D Ex ib IIIC T210°C Db
The input connections to the OPTIMASS 4010C / OPTIGAS 4010C for use with associated
apparatus have the following maximum values:
AD OPTIMASS 300_010 Haz Areas Rev 01.qxp 05/03/2014 13:24 Page 11

Other manuals for KROHNE OPTIMASS 8300C

Related product manuals